3M TRUE DEFINITION SCANNER 75030

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2015-11-25 for 3M TRUE DEFINITION SCANNER 75030 manufactured by 3m Espe Dental Products.

Event Text Entries

[32194950] Appropriate electrical safety measures (isolation transformer) are present in the true definition scanner system. The 3m true definition scanner instructions for use does contain the warning, 'to reduce the risks associated with hazardous voltage and fire - use only a properly grounded power outlet; do not use extension cords or multiple portable power socket outlets. '
Patient Sequence No: 1, Text Type: N, H10


[32194951] During a scanning procedure using a 3m true definition scanner, a patient experienced a shock and pain inside the mouth. It was confirmed that the power outlet, where the scanner was connected at the time the incident occurred, was not properly grounded. After the scanner was plugged into the power outlet where the ground was proven to be wired and functioning properly, no more shocks occurred. The dentist was advised of the investigation findings and agreed that the incident was caused by the faulty power outlet at the office; an electrician was called and fixed all power outlets in the office.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005174370-2015-00112
MDR Report Key5251988
Report SourceHEALTH PROFESSIONAL
Date Received2015-11-25
Date of Report2015-10-29
Date of Event2012-03-27
Date Mfgr Received2015-10-29
Date Added to Maude2015-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANGIE DRAPER
Manufacturer Street2510 CONWAY AVENUE
Manufacturer CityST. PAUL MN 551441000
Manufacturer CountryUS
Manufacturer Postal551441000
Manufacturer Phone6517331179
Manufacturer G13M ESPE DENTAL PRODUCTS
Manufacturer Street2510 CONWAY AVENUE
Manufacturer CityST. PAUL MN 551441000
Manufacturer CountryUS
Manufacturer Postal Code551441000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M TRUE DEFINITION SCANNER
Generic NameSYSTEM,OPTICAL IMPRESSION, COMPUTER ASSISTED DESIGN AND MANUFACTURING (CAD/CAM)
Product CodeNOF
Date Received2015-11-25
Catalog Number75030
Lot NumberG1525065
OperatorDENTIST
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M ESPE DENTAL PRODUCTS
Manufacturer Address2510 CONWAY AVENUE ST. PAUL MN 551441000 US 551441000


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-11-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.