VANTEX CVC WITH OLIGON MATERIAL ECVC3475

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-05-30 for VANTEX CVC WITH OLIGON MATERIAL ECVC3475 manufactured by Centurion Medical Products Corp..

Event Text Entries

[46570601] In preparation for catheter insertion, staff were attempting to flush the triple lumens of a central venous catheter and were unable to do so. After 5 separate catheter kits were tried and all were blocked to one degree or another, staff examined the end of the catheter and identified two teal cylindrical pieces of plastic approximately 1/4" long, one with a small piece of clear plastic attached and the other with the clear plastic piece separated from the teal piece. These teal pieces are the same color as the hub for the triple lumens. Both the medial and proximal lumens were blocked to some degree. These small pieces presented a serious patient safety issue and these catheters pulled from use and marked do not use. The next day, the decision was made to pull all of the manufacturer's catheters as a precaution and supply chain ordered replacement arrow catheters to use in the meantime.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5062582
MDR Report Key5694249
Date Received2016-05-30
Date of Report2016-05-30
Date of Event2016-05-24
Date Added to Maude2016-06-02
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Sequence Number: 0

Brand NameVANTEX CVC WITH OLIGON MATERIAL
Generic NameCENTRAL VENOUS CATHETER
Product CodeFOZ
Date Received2016-05-30
Model NumberECVC3475
Lot Number2016032250
Device Expiration Date2017-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No0
Device Event Key0
ManufacturerCENTURION MEDICAL PRODUCTS CORP.
Manufacturer AddressWILLIAMSTON MI 48895 US 48895

Device Sequence Number: 1

Brand NameVANTEX CVC WITH OLIGON MATERIAL
Generic NameCENTRAL VENOUS CATHETER
Product CodeLRS
Date Received2016-05-30
Model NumberECVC3475
Lot Number2016032250
Device Expiration Date2017-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerCENTURION MEDICAL PRODUCTS CORP.
Manufacturer AddressWILLIAMSTON MI 48895 US 48895


Patients

Patient NumberTreatmentOutcomeDate
10 2016-05-30

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.