POWERLOOK AMP WITH SECONDLOOK DIGITAL GE V7.2 DSC010-01-US

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-01-11 for POWERLOOK AMP WITH SECONDLOOK DIGITAL GE V7.2 DSC010-01-US manufactured by Icad, Incorporated.

Event Text Entries

[97380768] The oem field service engineer returned to customer site and installed the cad license and the system is working properly.
Patient Sequence No: 1, Text Type: N, H10


[97380769] A customer site indicated that no cad markings were reported from the cad after processing patient images. The oem mammography field service engineer (fse) did not apply the cad license after installing the cad software. This resulted in the site not having cad marks on cases for some time. No patient injury or adverse event was associated with this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1225671-2018-00001
MDR Report Key7183668
Date Received2018-01-11
Date of Report2018-01-11
Date of Event2017-12-14
Date Mfgr Received2017-12-14
Device Manufacturer Date2017-06-14
Date Added to Maude2018-01-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOHN DELUCIA
Manufacturer Street98 SPIT BROOK ROAD, SUITE 100
Manufacturer CityNASHUA NH 03062
Manufacturer CountryUS
Manufacturer Postal03062
Manufacturer Phone6033091945
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePOWERLOOK AMP WITH SECONDLOOK DIGITAL GE
Generic NameSECONDLOOK DIGITAL
Product CodeMYN
Date Received2018-01-11
Model NumberV7.2
Catalog NumberDSC010-01-US
OperatorRADIOLOGIC TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerICAD, INCORPORATED
Manufacturer Address98 SPIT BROOK ROAD, SUITE 100 NASHUA NH 03062 US 03062


Patients

Patient NumberTreatmentOutcomeDate
10 2018-01-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.