POLYETHYLENE EVA DISPOSABLE SMOOTH TUBING 243301

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2018-08-15 for POLYETHYLENE EVA DISPOSABLE SMOOTH TUBING 243301 manufactured by Tri-anim Health Services, Inc..

Event Text Entries

[117160416] Our respiratory therapists reported that they have had several occurrences of the 10 mm disposable smooth interior tubing "unraveling. " the thin membrane holding the coils together breaks easily with little flexion applied to the circuit, far less than would be under normal use conditions. Manufacturer response for ventilator tubing, (per site reporter). The device is manufactured by (b)(4), distributed by (b)(4) and repackaged by (b)(4). The lot numbers on the repackaged tubes we have are 579100 and 578973. After making a-m systems aware of the problems, (b)(4) contacted (b)(4) and has begun the recall process.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number7784207
MDR Report Key7784207
Date Received2018-08-15
Date of Report2018-08-08
Date of Event2018-07-28
Report Date2018-08-08
Date Reported to FDA2018-08-08
Date Reported to Mfgr2018-08-15
Date Added to Maude2018-08-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePOLYETHYLENE EVA DISPOSABLE SMOOTH TUBING
Generic NameTUBING, PRESSURE AND ACCESSORIES
Product CodeBYX
Date Received2018-08-15
Model Number243301
Catalog Number243301
Lot Number579100/578973
Device AvailabilityN
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerTRI-ANIM HEALTH SERVICES, INC.
Manufacturer Address5000 TUTTLE CROSSING BLVD. DUBLIN OH 43016 US 43016


Patients

Patient NumberTreatmentOutcomeDate
10 2018-08-15

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.