TROCAR HANDLE 25-472-05-07

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2018-10-03 for TROCAR HANDLE 25-472-05-07 manufactured by Karl Leibinger Medizintechnik Gmbh & Co. Kg.

Event Text Entries

[122533505] (b)(4). Multiple mdr reports were filed for this event, please see associated report: mdr: 9610905-2018-00236.
Patient Sequence No: 1, Text Type: N, H10


[122533506] It was reported a patient was burned while using a twist drill and trocar. The twist drill was being used at 20,000 rmps. It is stated in ifu (b)(4) "twist drills are not designed to be used at speeds in excess of 5000 rpm. "
Patient Sequence No: 1, Text Type: D, B5


[132430902] An investigation was performed on the basis of complaint statistics as no device was returned for evaluation. The complaint percentage falls within the design risk limits adhered to at klm. The review of the device history records was not possible due to no lot number being identified. The ifu indicates that drills are not to be used with than 5000 rmp. The provided information reveal that the root cause was an incorrect use by the customer. If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610905-2018-00237
MDR Report Key7931134
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2018-10-03
Date of Report2018-09-05
Date of Event2018-09-05
Date Facility Aware2018-12-18
Report Date2018-09-05
Date Reported to Mfgr2018-09-05
Date Mfgr Received2018-12-18
Date Added to Maude2018-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. JENNIFER DAMATO
Manufacturer StreetP.O. BOX 16369
Manufacturer CityJACKSONVILLE FL 32245
Manufacturer CountryUS
Manufacturer Postal32245
Manufacturer Phone9046417746
Manufacturer G1KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
Manufacturer StreetKOLBINGER STRASSE 10
Manufacturer CityMUEHLHEIM/DONAU, 78570
Manufacturer CountryGM
Manufacturer Postal Code78570
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTROCAR HANDLE
Generic NameTROCAR HANDLE
Product CodeLRC
Date Received2018-10-03
Model Number25-472-05-07
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerKARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
Manufacturer AddressKOLBINGER STRASSE 10 MUEHLHEIM/DONAU, 78570 GM 78570


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2018-10-03

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.