SOUTHMEDIC #11 NON-STERILE DISPOSABLE SCALPEL 73-0211

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor,health profession report with the FDA on 2020-02-25 for SOUTHMEDIC #11 NON-STERILE DISPOSABLE SCALPEL 73-0211 manufactured by Southmedic Inc..

Event Text Entries

[180807108] The manufacturer was notified on 12/13/2019 by the distributor of a complaint against scalpel 73-0211. They advised that their alert date was (b)(6) 2019 and they have not yet received samples. Should samples be received from the distributor, southmedic shall evaluate them for further investigation. Review of batch records demonstrate that part is within specifications. The submission delay of this report is explained in attachment 1.
Patient Sequence No: 1, Text Type: N, H10


[180807109] As medical resident went to remove clear pvc cover to scalpel, it was stiff and required extra force and during removal, sliced his fingers. Resident very cautious to use kit now and hospital now providing a swanmorton single use scalpel - which does not have a cover over it.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8022032-2019-00005
MDR Report Key9750917
Report SourceDISTRIBUTOR,HEALTH PROFESSION
Date Received2020-02-25
Date of Report2020-01-03
Date of Event2019-11-26
Date Mfgr Received2019-12-13
Device Manufacturer Date2018-10-29
Date Added to Maude2020-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactTISH WHITEHEAD
Manufacturer Street50 ALLIANCE BLVD.
Manufacturer CityBARRIE, ONTARIO L4M 5K3
Manufacturer CountryCA
Manufacturer PostalL4M 5K3
Manufacturer G1SOUTHMEDIC INC.
Manufacturer Street50 ALLIANCE BLVD.
Manufacturer CityBARRIE, L4M 5K3
Manufacturer CountryCA
Manufacturer Postal CodeL4M 5K3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSOUTHMEDIC #11 NON-STERILE DISPOSABLE SCALPEL
Generic NameSCALPEL
Product CodeGES
Date Received2020-02-25
Model Number73-0211
Catalog Number73-0211
Lot Number102918
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSOUTHMEDIC INC.
Manufacturer Address50 ALLIANCE BLVD BARRIE, L4M 5K3 CA L4M 5K3


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2020-02-25

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.