MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-02-20 for DRIED POS MIC TYPE 20A B1017-203A manufactured by Siemens Healthcare Diagnostics Inc.
[801478]
Customer reported s aureus isolates oxacillin (ox) mic discrepancy. They obtained oxacillin-susceptible results on the dried pos mic type 20a panel and oxacillin-resistant results on secondary methods that were also performed for the clinical isolate. It is unk if the results were reported to the physician. No reports of adverse health consequences associated with the discrepant result being obtained. The cause of the oxacillin susceptible results is unk.
Patient Sequence No: 1, Text Type: D, B5
[8088020]
Evaluation, - routine monitoring of complaint history and performance trends to ensure performance is within claims. Evaluation results - requested additional information to determine if malfunction occurred. Evaluation conclusions - product is within performance claims. The cause of the oxacillin susceptible results is unk.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2919016-2008-00003 |
MDR Report Key | 1002088 |
Report Source | 05 |
Date Received | 2008-02-20 |
Date of Report | 2008-02-12 |
Date of Event | 2008-01-28 |
Date Mfgr Received | 2008-01-28 |
Device Manufacturer Date | 2007-11-01 |
Date Added to Maude | 2008-05-06 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | RICH LAGUNA |
Manufacturer Street | 2040 ENTERPRISE BLVD. |
Manufacturer City | WEST SACRAMENTO CA 95691 |
Manufacturer Country | US |
Manufacturer Postal | 95691 |
Manufacturer Phone | 9163743163 |
Single Use | 3 |
Previous Use Code | 3 |
Removal Correction Number | NA |
Event Type | 3 |
Type of Report | 3 |
Brand Name | DRIED POS MIC TYPE 20A |
Generic Name | ANTIMICROBIC SUSCEPT. PANEL: GRAM POS |
Product Code | LTT |
Date Received | 2008-02-20 |
Model Number | NA |
Catalog Number | B1017-203A |
Lot Number | 2008-11-06 |
ID Number | NA |
Device Expiration Date | 2008-11-06 |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | NA |
Device Eval'ed by Mfgr | R |
Implant Flag | N |
Date Removed | B |
Device Sequence No | 1 |
Device Event Key | 999318 |
Manufacturer | SIEMENS HEALTHCARE DIAGNOSTICS INC |
Manufacturer Address | 2040 ENTERPRISE BLVD. WEST SACRAMENTO CA 95691 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2008-02-20 |