HENRY SCHEIN ONE STEP + HCG DIPSTICK FHC-101-KHS25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2008-03-04 for HENRY SCHEIN ONE STEP + HCG DIPSTICK FHC-101-KHS25 manufactured by Innovacon, Inc..

Event Text Entries

[784176] In 2008, a woman did an ept test and the result was positive. Lauren had this woman collect a urine specimen around 2 pm and lauren tested this specimen on henry schein urine hcg rapid dipstick test. Result was negative. The woman went to her ob/gyn that same day around 6 pm where they did a urine test (result positive) and a serum test (result positive). It was unknown if it was a serum qualitative or quantitative that was performed. Qc is working fine and seeing a control line with each...
Patient Sequence No: 1, Text Type: D, B5


[7951264] Retention device testing performed. The retention devices meet qc specification. Correct positive results were observed when tested with 25 miu/ml hcg urine control. The 100 miu/ml 10 iu/ml and 213. 2 iu/ml hcg urine control yielded clearly positive results at 3 minutes reading time. Probable root cause: the urine hcg level always lower than serum as the urine sample is more likely influenced by many factors. If the patient drinks too much water which will dilute the urine before the test, the urine may not contain representative levels of hcg and a false negative results may be obtained. As so, if pregnancy is suspected, a first morning urine specimen should be collected and tested. Conclusion: this complaint is not confirmed. Manufacturing documentation for this lot number was review. No abnormality had been observed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005689981-2008-00013
MDR Report Key1006618
Report Source05,08
Date Received2008-03-04
Date of Report2008-02-29
Date of Event2008-01-11
Date Mfgr Received2008-02-07
Date Added to Maude2008-05-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street4106 SORRENTO VALLEY BLVD.
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8582008910
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHENRY SCHEIN ONE STEP + HCG DIPSTICK
Generic NameONE STEP PREGNANCY TEST (URINE)
Product CodeJHJ
Date Received2008-03-04
Model NumberFHC-101-KHS25
Lot NumberHCG7090098
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1005451
ManufacturerINNOVACON, INC.
Manufacturer Address4106 SORRENTO VALLEY BLVD. SAN DIEGO CA 92121 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-03-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.