AXSYM DRUGS FOR ABUSE TOXICOLOGY ASSAY DISK 3D54-08

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2008-03-06 for AXSYM DRUGS FOR ABUSE TOXICOLOGY ASSAY DISK 3D54-08 manufactured by Abbott Laboratories.

Event Text Entries

[807095] The customer stated that vrtx error 002 in task 40 occurred on the axsym analyzer after editing the cutoff parameters for amphetamines. Axsym system software version 8. 0 was installed on the analyzer. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


[7945933] A new axsym drugs of abuse (doa)/toxicology assay disk (version 8. 0) was released containing an updated version of the axsym amphetamine/methamphetamine ii assay file. This file was created using axsym offline software version 6. 0. When customers attempted to edit the axsym amphetamine/ meth-amphetamine ii positive and negative interpretation cutoffs after installing the updated assay file, they observed an axsym system vrtx error #002 in task 40 when running patient samples. The vrtx error causes the instrument to lock up. Through the analysis of existing data and the preliminary investigation associated with the risk evaluation, the root cause was identified to be a change in the axsym offline software. Axsym offline software version 6. 0 replaced version 5. 9. 0. The parameters in axsym offline software version 5. 9. 0 contained 31 characters while the parameters in axsym offline software version 6. 0 contained 47. The offline software pads these extra fields with spaces. When the customer edits the positive and negative interpretation cutoffs the characters from these fields are pushed into the next fields, the negative recalculation interpretation formula and the positive interpretation hold option parameters. When an amphetamine result is generated, the axsym system software attempts to interpret the result. Because space characters have been pushed from the positive and negative interpretation cutoff parameters into the interpretation formula and hold option parameters, the axsym software cannot make a decision and the vrtx error is generated. On 01/10/08, the doa/toxicology assay disk was put on quality hold. A new disk will be released to the field using axsym offline software version 5. 9. 0. No later than 3/31/08, which will have addressed the issue in axsym offline software version 6. 0. This is the final report. End of report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2008-00037
MDR Report Key1008571
Report Source01
Date Received2008-03-06
Date of Report2008-02-19
Date of Event2008-02-19
Date Mfgr Received2008-02-19
Date Added to Maude2008-06-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSUDARSHAN HEBBAR, MD, D-3MQC, AP6B
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 600646112
Manufacturer CountryUS
Manufacturer Postal600646112
Manufacturer Phone8479379677
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction Number1415939-2/01/08-001-C
Event Type3
Type of Report3

Device Details

Brand NameAXSYM DRUGS FOR ABUSE TOXICOLOGY ASSAY DISK
Generic NameAMPHETAMINE TEST SYSTEM
Product CodeMGX
Date Received2008-03-06
Model NumberNA
Catalog Number3D54-08
Lot NumberNA
ID NumberVERSION 8.0
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1005290
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 600643500 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-03-06