REPROCESSED ETHICON TROCAR XCEL BLUNT TIP H12LP

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-02-07 for REPROCESSED ETHICON TROCAR XCEL BLUNT TIP H12LP manufactured by Ascent Healthcare Solutions.

Event Text Entries

[808206] A total of 5 reprocessed trocars broke apart during use. One was from lot number 149680, three were from lot number 149764, and one was from an unknown lot number. The five broken trocars were sent to the manufacturer (reprocessor) for analysis. The reprocessor's analysis indicated that, based on their findings, a change would be made in the disassembly procedure involved in reprocessing this trocar. According to the reprocessor, "this modification will assist in mitigating future occurrences of this nature. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1009819
MDR Report Key1009819
Date Received2008-02-07
Date of Report2008-01-30
Date of Event2008-01-21
Report Date2008-01-30
Date Reported to FDA2008-02-07
Date Added to Maude2008-03-12
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameREPROCESSED ETHICON TROCAR
Generic NameTROCAR
Product CodeNMK
Date Received2008-02-07
Returned To Mfg2007-12-03
Model NumberXCEL BLUNT TIP
Catalog NumberH12LP
Lot Number149680
ID Number*
OperatorPHYSICIAN
Device AvailabilityR
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key977338
ManufacturerASCENT HEALTHCARE SOLUTIONS
Manufacturer Address10232 SOUTH 51ST STREET PHOENIX AZ 850445205 US

Device Sequence Number: 2

Brand NameREPROCESSED ETHICON TROCAR
Generic NameTROCAR
Product CodeNMK
Date Received2008-02-07
Returned To Mfg2008-12-01
Model NumberH12LP
Catalog Number*
Lot Number149764
ID Number*
Device AvailabilityR
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key977339
ManufacturerASCENT HEALTHCARE SOLUTIONS
Manufacturer Address10232 S 51ST STREET PHOENIX AZ 850445205 US

Device Sequence Number: 3

Brand NameREPROCESSED ETHICON TROCAR
Generic NameTROCAR
Product CodeNMK
Date Received2008-02-07
Returned To Mfg2008-12-01
Model NumberXCEL BLUNT TIP
Catalog Number*
Lot NumberUNK
ID Number*
Device AvailabilityR
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No3
Device Event Key977340
ManufacturerASCENT HEALTHCARE SOLUTIONS
Manufacturer Address10232 S 51ST STREET PHOENIX AZ 850445205 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-02-07