BBL THIOGLYCOLLATE MEDIUM, ENRICHED WITH VIT 297292

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-03-11 for BBL THIOGLYCOLLATE MEDIUM, ENRICHED WITH VIT 297292 manufactured by Bd Diagnostics.

Event Text Entries

[837172] Customer called to report that about half of her box of 100 tubes were too tight and hard to get open. One of her lab assistance had a tube break in her hands upon trying to open it and cut her thumb requiring stitches.
Patient Sequence No: 1, Text Type: D, B5


[8029343] Seven tubes from this lot were returned from the customer for investigation. Two tubes were opened by hand; one was easy to pen and the other was difficult. The remaining five tubes were tested for torque with readings from 4. 6 to 7. 6 in/lb. The satisfactory range is 3. 0-8. 0 in/lb with a target of 5. 5 in/lb. Additionally, the difficult to open tubes were measured for critical dimensions and found to be within spec. Retention samples were tested for ease to open and torque. Tubes were easily opened by a lab tech and the torque testing results were between 1. 3 to 5. 7 in/lb. The batch history record was examined and found to be satisfactory with all torque readings between 4. 5 and 6. 4 in/lb during filling. A review of the complaint sys was conducted and show no other reported of this defect on this prod lot. This lot will be monitored for future reports of this defect. A corrective action has been opened to further address this issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1119779-2008-00004
MDR Report Key1011985
Report Source05
Date Received2008-03-11
Date of Report2008-03-10
Date of Event2008-02-12
Date Mfgr Received2008-02-12
Device Manufacturer Date2007-11-01
Date Added to Maude2008-08-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN GERLICH
Manufacturer Street7 LOVETON CIR.
Manufacturer CitySPARKS MD 211520999
Manufacturer CountryUS
Manufacturer Postal211520999
Manufacturer Phone4103164464
Manufacturer G1BECTON DICKINSON
Manufacturer Street250 SCHILLIBNG CIR
Manufacturer CityCOCKEYSVILLE MD 21030
Manufacturer CountryUS
Manufacturer Postal Code21030
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBBL THIOGLYCOLLATE MEDIUM, ENRICHED WITH VIT
Product CodeJSG
Date Received2008-03-11
Returned To Mfg2008-02-18
Catalog Number297292
Lot Number7323024
Device Expiration Date2008-11-11
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1008540
ManufacturerBD DIAGNOSTICS