EXH VLV ASSY, PED, MILL, ENGLISH 21442-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,08 report with the FDA on 2008-03-13 for EXH VLV ASSY, PED, MILL, ENGLISH 21442-1 manufactured by Sechrist Industries, Inc..

Event Text Entries

[15159279] Distributor in (b) (4) reported that customer at hospital apparently attempted to use ventilator with reference pressure line stuck into a threaded hole in the front of the pediatric exhalation block. The pediatric exhalation block is used to ventilate pts ranging from 10 lbs to 100 lbs, not neonates or premature infants. The threaded hole is plugged with at least two setscrews to prevent full insertion of the sense line. The reference pressure connector is designed to be a tapered slip fit into a smooth side tapered hole in the lower front of the block. The block label clearly identifies the holes the proximal line, reference pressure line and "from pt" hose connections are to be made into.
Patient Sequence No: 1, Text Type: D, B5


[15172702] Appears to be user error. Follow up report will be submitted for evaluation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2020676-2008-00003
MDR Report Key1016373
Report Source01,06,08
Date Received2008-03-13
Date of Report2008-03-13
Date of Event2008-02-04
Date Facility Aware2008-02-18
Date Mfgr Received2008-02-12
Device Manufacturer Date2000-09-01
Date Added to Maude2009-12-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactGREG GODFREY
Manufacturer Street4225 EAST LA PALMA AVE.
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal92807
Manufacturer Phone7145798400
Manufacturer G1SECHRIST INDUSTRIES, INC.
Manufacturer Street4225 EAST LA PALMA AVE.
Manufacturer CityANAHEIM CA 92807
Manufacturer CountryUS
Manufacturer Postal Code92807
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEXH VLV ASSY, PED, MILL, ENGLISH
Generic NamePEDIATRIC EXHALATION VALVE
Product CodeCBP
Date Received2008-03-13
Model Number21442-1
Catalog Number21442-1
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSECHRIST INDUSTRIES, INC.
Manufacturer AddressANAHEIM CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-03-13