CELLIENT 71663-001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2008-03-26 for CELLIENT 71663-001 manufactured by Cytyc Corp..

Event Text Entries

[20238633] Customer was not satisfied with cellularity of cellient blocks with known tumor. Upon discussion, the physician pointed out his concerns and disappointment with our equipment. He said that when he has a sample with a lot of material he has problems in obtaining good results with cellient. He also expressed concern because usually the type of pts that come with this sample are the ones with a had clinical picture meaning that the probability of finding cancer is extremely high. He said that our filter did not have the capability to process enough material to make him comfortable or make a diagnosis on any pt that has a serious tumor. He ran a cell block through cellient on different occasions and then used the remainder of the material to run a normal cell block. The results are better using the normal procedure rather than when using cellient. He feels that the cellient equipment is great for thyroid fna or for any applications that have little material on it. The sample he ran on the equipment was liquid. He also showed three to five slides prepared on cellient where the sample tissue after the process did not come as a full circle on the slide. It has less than a full circular formation and others look like a half moon.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2008-00033
MDR Report Key1019630
Report Source06
Date Received2008-03-26
Date of Report2008-03-26
Date of Event2008-02-28
Date Mfgr Received2008-02-28
Date Added to Maude2008-07-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRAJ KASBEKAR, MGR
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5082638858
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCELLIENT
Generic NameCELLIENT
Product CodeKET
Date Received2008-03-26
Model Number71663-001
Catalog NumberNI
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1018680
ManufacturerCYTYC CORP.
Manufacturer AddressMARLBOROUGH MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-03-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.