COBAS AMPLIPREP/COBAS TAGMAN HIV-1 03542998190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2008-04-03 for COBAS AMPLIPREP/COBAS TAGMAN HIV-1 03542998190 manufactured by Roche Molecular Systems, Inc..

Event Text Entries

[815346] A patient undergoing therapy was monitored for using the suspect medical device. Results generated for three months in 2007 and in 2008 were 307 copies/ml, 491 copied/ml, target not detected (tnd) and tnd, respectively. Testing in 2008 also yielded tnd results. Because cd4 count was dropping, the treating physician requested additional testing using another test. Bdna results indicate a viral load of 117, 248 copies/ml. Testing the following month on a new specimen collection generated results of tnd for the cobas ampliprep/cobas tagman test and 14,398 copies/ml by bdna test.
Patient Sequence No: 1, Text Type: D, B5


[8026919] It is unknown when patient treatment was initiated or if it was altered based on these results. An investigation is in progress to get more details regarding the incident. Sample material has been identified to be sent for sequencing. 2008.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2243471-2008-00005
MDR Report Key1024395
Report Source05,07
Date Received2008-04-03
Date of Report2008-04-03
Date of Event2008-02-28
Date Mfgr Received2008-03-20
Date Added to Maude2008-04-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactROBERT PIGOZZI
Manufacturer Street4300 HACIENDA DRIVE
Manufacturer CityPLEASANTON CA 94588
Manufacturer CountryUS
Manufacturer Postal94588
Manufacturer Phone9257308272
Manufacturer G1ROCHE MOLECULAR SYSTEMS, INC.
Manufacturer Street1080 U.S. HIGHWAY 202 BRANCHBURG TOWNSHIP
Manufacturer CitySOMERVILLE NJ 08876377
Manufacturer CountryUS
Manufacturer Postal Code08876 3771
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOBAS AMPLIPREP/COBAS TAGMAN HIV-1
Generic NameHIV-1 NUCLEIC ACID TEST
Product CodeMTL
Date Received2008-04-03
Catalog Number03542998190
Lot NumberJ04456
Device Expiration Date2008-03-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key992064
ManufacturerROCHE MOLECULAR SYSTEMS, INC.
Manufacturer AddressBRANCHBURG NJ 08876 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-04-03

© 2026 FDA.report