PROVOX VOICE PROSTHESIS REF 7202

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,08 report with the FDA on 2008-04-02 for PROVOX VOICE PROSTHESIS REF 7202 manufactured by Atos Medical Ab.

Event Text Entries

[827370] According to the complaint: hospital found a hole in plastic cover, and the sterility was therefore ruined. They suspect that the scalpel inside has caused the hole.
Patient Sequence No: 1, Text Type: D, B5


[7994236] Visual examination of the device packaging shows compression damage on both ends of the outer package indicating very rough handling of the package. The inner package is mainly intact apart from a hole at the center were the tip of the scalpel has penetrated the plastic container and a dent inwards on the package at the handle end of the scalpel. Postmarket surveillance shows that the current design of the package has been used since 1995 and large quanity of the devices have been distributed world wide. No previous complaints indicating any packaging is suitable for its intended use. The investigation has concluded that accidental and unforeseen shipping and/or handling damage of the outer carton has caused the damage to the device packaging. A caution on the packaging states that the product is sterile if the package is undamaged and since no previous complaints have been reported no further actions are taken.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8032044-2008-00002
MDR Report Key1025316
Report Source05,08
Date Received2008-04-02
Date of Report2008-03-25
Date of Event2008-03-03
Date Mfgr Received2008-03-03
Device Manufacturer Date2005-02-28
Date Added to Maude2009-01-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFERENC DAHNER
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY
Manufacturer CountrySW
Manufacturer Phone41519800
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROVOX VOICE PROSTHESIS
Generic NameVOICE PROSTHESIS
Product CodeESE
Date Received2008-04-02
Returned To Mfg2008-03-12
Model NumberPROVOX
Catalog NumberREF 7202
Lot Number0502004
Device Expiration Date2010-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressPO BOX 183 HORBY SE-242 2 SW SE-242 22


Patients

Patient NumberTreatmentOutcomeDate
10 2008-04-02

© 2024 FDA.report