5943API ADVANCED CONTROL PADS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2008-04-04 for 5943API ADVANCED CONTROL PADS manufactured by Mizuho Orthopedic Systems, Inc..

Event Text Entries

[17773360] After surgery (8+hrs) the pt's skin was breaking down.
Patient Sequence No: 1, Text Type: D, B5


[17963870] Investigation found the end-user inappropriately positioning the placement and using the wrong pads. They were also inappropriately positioning the pt on the pads and table. It was also determined that the hospital was not using the pt care kit to reduce abrasions and stiction of the pts on the pads. An inservice with the hosp has been requested.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2921578-2008-00007
MDR Report Key1025913
Report Source01
Date Received2008-04-04
Date of Report2008-03-01
Date of Event2008-02-19
Date Mfgr Received2008-03-06
Date Added to Maude2010-03-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKIRKE JAYNE
Manufacturer Street30031 AHERN AVE.
Manufacturer CityUNION CITY CA 945871234
Manufacturer CountryUS
Manufacturer Postal945871234
Manufacturer Phone5104768128
Manufacturer G1MIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Street30031 AHERN AVE.
Manufacturer CityUNION CITY CA 94587123
Manufacturer CountryUS
Manufacturer Postal Code94587 1234
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name5943API ADVANCED CONTROL PADS
Generic NameMODULAR OPERATING ROOM TABLE
Product CodeFWY
Date Received2008-04-04
Model Number5943API
Catalog Number5943API
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMIZUHO ORTHOPEDIC SYSTEMS, INC.
Manufacturer Address30031 AHERN AVE. UNION CITY CA 94587123 US 94587 1234


Patients

Patient NumberTreatmentOutcomeDate
158031. Other 2008-04-04

© 2026 FDA.report