INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004051

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2008-04-07 for INDEPENDENCE IBOT 4000 MOBILITY SYSTEM IT004051 manufactured by Independence Technology, Llc..

Event Text Entries

[19957705] A user reported a forward fall in the device while being assisted to descend a set of stairs in assisted stair climbing function. The reported fall resulted in a broken nose that required medical attention. The user stated that instead of leaning backward during the descent, he leaned forward, which caused the device to fall forward and down the steps on top of him. No product malfunction is indicated. This report corresponds to independence technology complaint.
Patient Sequence No: 1, Text Type: D, B5


[20116733] In either solo or assisted stair climbing function, users are an active component of the stair climb function, and are trained to lean in the correct direction in order to facilitate the completion of the ascent or descent. Service was dispatched to retrieve the electronic configuration file (ecf) in 2008. A report on field service activity (sar) and a device checkout record (fcr) will be forwarded to the complaint handling unit (chu) per standard operating procedure. Company review of the ecf confirms that the pitch of the device was in the forward direction, and with additional forward pitch, the cluster rotated. The device pitch was adequately corrected to slow the rotation, however, as the cluster rotation completed, the device was aggressively pitched forward again, most likely due to the forward lean described by the user. Device immediately entered cluster safety lock and continued to pitch forward. The assistant did not respond properly by preventing the device from continuing to pitch forward. Cluster safety lock is the device's response to an out of control stair climbing condition. The user has not reported any recurrence of the described event since the completion of the service activity.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3003508375-2008-00002
MDR Report Key1026099
Report Source04
Date Received2008-04-07
Date of Report2008-03-31
Date of Event2008-03-20
Date Mfgr Received2008-03-20
Device Manufacturer Date2007-11-01
Date Added to Maude2009-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHAEL O'MEARA, DIRECTOR
Manufacturer Street45 TECHNOLOGY DRIVE
Manufacturer CityWARREN NJ 07059
Manufacturer CountryUS
Manufacturer Postal07059
Manufacturer Phone9084122200
Manufacturer G1CREATIVE TECHNOLOGY SERVICE
Manufacturer Street7444 HAGGERTY ROAD
Manufacturer CityCANTON MI 48187
Manufacturer CountryUS
Manufacturer Postal Code48187
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINDEPENDENCE IBOT 4000 MOBILITY SYSTEM
Generic NameSTAIR CLIMBING WHEELCHAIR
Product CodeIMK
Date Received2008-04-07
Model NumberNA
Catalog NumberIT004051
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINDEPENDENCE TECHNOLOGY, LLC.
Manufacturer Address45 TECHNOLOGY DRIVE WARREN NJ 07059 US 07059


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2008-04-07