TAO BRUSH I.U.M.C ENDOMETRIAL SAMPLER J-ES-090500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2008-05-01 for TAO BRUSH I.U.M.C ENDOMETRIAL SAMPLER J-ES-090500 manufactured by Cook Urological, Inc..

Event Text Entries

[853573] Customer states: "the ball tip of the tao brush had come off inside the patient's uterus, and they knew the tip was on the brush prior to placement. "
Patient Sequence No: 1, Text Type: D, B5


[7998917] One specimen jar containing the brush portion in fluid and the brush shaft was received. The clear sheath on the shaft was noted to be slid up 3cm from the handle in the open position. The clear sheath was also noted to be separated into two segments with the point of separation to have a twisted appearance. The shaft segment measured 19. 4cm and the brush segment measured 2. 9cm. The wires were noted to be cut below the bristles of the brush; cut marks were visible on the wires on both segments. Cutting the brush off following the procedure is common practice. Upon close examination of the distal end, where the ball tip was located, adhesive was present. The distributor indicated the physician and patient agreed to not remove the ball at the time of the separation; the physician indicated the physician and patient agreed to not remove the ball at the time of the separation; the physician indicated the ball should come out on it's own. The ball tip is made of polythene plastic and is approximately 3mm in diameter. The distributor has been contacted for additional information and has indicated they have not been made aware as to whether the patient has passed the ball tip or not. They also indicated that they were not getting any additional information from the physician. Each brush has the ball tip 100% pull tested during the manufacturing phase to assure the ball is secure prior to packaging. We are unable to determine what has caused the ball to separate from the shaft of the brush. Should additional information become available, it will be forwarded.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1825146-2008-00018
MDR Report Key1036638
Report Source08
Date Received2008-05-01
Date Mfgr Received2008-04-08
Device Manufacturer Date2006-11-01
Date Added to Maude2008-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactTAMMY BACON
Manufacturer Street1100 WEST MORGAN ST.
Manufacturer CitySPENCER IN 47460
Manufacturer CountryUS
Manufacturer Postal47460
Manufacturer Phone8128294891
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTAO BRUSH I.U.M.C ENDOMETRIAL SAMPLER
Generic NameTAO BRUSH
Product CodeHHK
Date Received2008-05-01
Returned To Mfg2008-04-14
Model NumberNA
Catalog NumberJ-ES-090500
Lot NumberU1586825
ID NumberNA
Device Expiration Date2009-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1005737
ManufacturerCOOK UROLOGICAL, INC.
Manufacturer AddressSPENCER IN 47460 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-05-01