TCM II Y CIRCUIT 14191010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1993-12-06 for TCM II Y CIRCUIT 14191010 manufactured by Sarns 3m Health Care, Inc..

Event Text Entries

[5298] During cardiopulmonary bypass, following removal of the aortic clamp, the h20 temperature of the systemic heater/cooler was allowed to cool rather than warm. The water temperature dropped from 40 degrees c to 26 degrees c. This caused a drip in blood temperature from 36 degress c to 28 degrees c. A different heater/cooler was brought in to complete the procedure. No adverse effect on pt. A combination of valves and tees recommended by the mfr may contribute to operator error.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1000016
MDR Report Key10377
Date Received1993-12-06
Date of Report1993-11-12
Date of Event1993-09-16
Date Added to Maude1993-12-07
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameTCM II Y CIRCUIT
Generic Name1/2" 2-WAY BALL VALVE
Product CodeMJJ
Date Received1993-12-06
Model Number14191010
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key10377
ManufacturerSARNS 3M HEALTH CARE, INC.
Manufacturer AddressANN ARBOR MI 48103 US


Patients

Patient NumberTreatmentOutcomeDate
10 1993-12-06

© 2026 FDA.report