FRONTAL SINUS MINI-TREPHINE SET 1892000

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-05-14 for FRONTAL SINUS MINI-TREPHINE SET 1892000 manufactured by Medtronic Xomed, Inc..

Event Text Entries

[849012] Event occurred in pt being treated for severe afs (allergic fungal sinusitis) and small left orbital mucocele. Irrigation through the minitrephine was initially uneventful with fungal debris movement and the fluorescein stained saline moving normally. After 20cc irrigation placed and irrigated, there was no flow through the nose. Noted left eye bulging, did a lateral canthotomy/cantholysis and medial decompression. Suspect a large blood of fungal debris blocked the irrigation outflow tract, and path of least resistance became the intraorbital route. Orbital exploration showed fluorescein stained saline in supero-medial compartment, ct showed supero-lateral fluid collection in orbit.
Patient Sequence No: 1, Text Type: D, B5


[8018405] Event was described in discussions with the doctor. Abnormal or unusual variation in pt's anatomy caused the event and, therefore, it was not necessary to evaluate the device. Successful revision procedure was performed three days later. At that time, the doctor reported that the pt's vision was nearly normal.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2008-00009
MDR Report Key1045336
Report Source05
Date Received2008-05-14
Date of Report2008-05-14
Date of Event2008-04-14
Date Mfgr Received2008-04-15
Date Added to Maude2008-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMIKE MOSBY
Manufacturer Street6743 SOUTHPOINT DR. NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9042797584
Manufacturer G1MEDTRONIC XOMED
Manufacturer Street6743 SOUTHPOINT DR. NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFRONTAL SINUS MINI-TREPHINE SET
Generic NameSINUS TREPHINATION INSTRUMENT SET
Product CodeKBF
Date Received2008-05-14
Model Number1892000
Catalog Number1892000
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1015239
ManufacturerMEDTRONIC XOMED, INC.
Manufacturer AddressJACKSONVILLE FL 32216 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2008-05-14