ADVIA CHEMISTRY TRIGLYCERIDE METHOD B01-4133-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2008-05-24 for ADVIA CHEMISTRY TRIGLYCERIDE METHOD B01-4133-01 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[849347] Customer reported a discordant triglyceride result when using reagent lot # 095939. The sample initially read 483 mg/dl and upon retest with a diluted sample, the corrected value was 2120 mg/dl. The customer became aware of this problem when she noticed that the reaction curve for the initial result was discrepant. There was no patient intervention as a result of this discordant triglyceride result.
Patient Sequence No: 1, Text Type: D, B5


[8079479] Discussions with the oem reagent manufacturer indicates that the advia chemistry triglyceride method kit lot 095939 is exhibiting lower than expected linearity. The root cause. Investigation is under way by the manufacturer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2008-00060
MDR Report Key1051308
Report Source05,06
Date Received2008-05-24
Date of Report2008-03-14
Date of Event2008-02-25
Date Mfgr Received2008-03-14
Date Added to Maude2009-05-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPETRA KANDUS
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145243392
Manufacturer G1RANDOX LABORATORIES LTD
Manufacturer CityCRUMLIN, COUNTY ANTRIM BT294QY
Manufacturer CountryEI
Manufacturer Postal CodeBT294QY
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVIA CHEMISTRY TRIGLYCERIDE METHOD
Generic NameTRIGLYCERIDE METHOD
Product CodeJGY
Date Received2008-05-24
Model NumberNA
Catalog NumberB01-4133-01
Lot NumberRGT. LOT # 095939
ID NumberNA
Device Expiration Date2009-01-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2008-05-24

© 2026 FDA.report