FLEXIBLE TIP WITH SLIMLINE RETENTION 119042

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-05-16 for FLEXIBLE TIP WITH SLIMLINE RETENTION 119042 manufactured by Covidien.

Event Text Entries

[18332736] Customer reports enema tip separated from the clear plastic adapter that attaches to the bag and entered the patient's body during a procedure. Patient was a male. Complainant inspected other tips in the box and found at least one other where the tip were not securely fastened to the adapter. It slides out as if the adapter was too small or there was no glue on it (not snug). Two cases will be returned for investigation. The used tips were discarded.
Patient Sequence No: 1, Text Type: D, B5


[18346527] Root cause: this defect is an operational error due that the glue is applied to the tip and adapter manually. Conclusions: device history record was reviewed finding no defects detected in the inspection of this lot. Twenty one samples were received from lot number 3202055 of model number 119042. All samples were evaluated under procedure for tensile strength and in fact, in one of the samples, the plastic adapter gets loose. Therefore this complaint was confirmed. Corrective actions: a notification report is being issued to inform all personnel within the enema tips assembly process about this problem. Our production process is being checked putting special attention on this issue. With the reference reported, notification was provided to appropriate manufacturing and quality assurance personnel.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610849-2008-00010
MDR Report Key1051373
Report Source05
Date Received2008-05-16
Date of Report2003-12-04
Date of Event2003-12-04
Date Mfgr Received2003-12-04
Device Manufacturer Date2006-07-01
Date Added to Maude2008-09-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID BENSON
Manufacturer Street211 EAST GALBRAITH ROAD
Manufacturer CityCINCINNATI OH 45237
Manufacturer CountryUS
Manufacturer Postal45237
Manufacturer Phone5139485719
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameFLEXIBLE TIP WITH SLIMLINE RETENTION
Generic NameENEMA TIP
Product CodeFCE
Date Received2008-05-16
Model NumberFLEXIBLE TIP
Catalog Number119042
Lot Number3202055
ID NumberNA
Device Expiration Date2006-07-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1045718
ManufacturerCOVIDIEN