OLYMPUS WA50011A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2008-05-23 for OLYMPUS WA50011A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[863712] The user facility reported that during a therapeutic laparoscopy, a complete loss of image was experienced. The procedure was reportedly completed with a different, but similar device and there was no patient injury. No further information was provided by the user facility.
Patient Sequence No: 1, Text Type: D, B5


[8085749] The device referenced in this report was returned to olympus for investigation. The investigation did not duplicate the user's report of a complete loss of image. However, the image was observed to be slightly dark, and the unit failed the light transmission test. The investigation noted large dents on the outer tube. Additionally, dents were noted on the light guide adapter and video connector plug and scratches were found on the outer tube, distal end, and video connector plug. The cause of the user's experience cannot be conclusively determined, however, the physical damage found on the device cannot be ruled out as a contributing factor to the reported event. This report is being filed as medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2008-00016
MDR Report Key1052379
Report Source06
Date Received2008-05-23
Date of Report2008-04-24
Date of Event2008-04-24
Date Added to Maude2010-04-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameVIDEO TELESCOPE
Product CodeFBP
Date Received2008-05-23
Model NumberWA50011A
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressHAMBURG


Patients

Patient NumberTreatmentOutcomeDate
10 2008-05-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.