MAUDE MDR 1054098

MDR report key
1054098
Report number
2431026-2008-00003
Event key
0
Event type
3
Date of event
2008-04-25
Date received
2008-06-02
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
0

Manufacturer Contact#

Contact
KENT JONES
Address
22 CARL ZEISS STRASSE OBERKOCHEN D-734 GM
Phone
925-925-9255
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1OPMI PENTEROSURGICAL MICROSCOPECARL ZEISS SURGICAL GMBHEPT1290-983NYY

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12008-06-0201. O

Event Narratives#

D

Patient 1

THE USER FACILITY REPORTED AN INCIDENT DESCRIBED AS A PT BURN ("LEVEL 2"). THE PT EXHIBITED A CIRCULAR BURN SURROUNDING THE SURGICAL AREA WHERE AN INCISION HAD BEEN MADE TO CONDUCT A SKIN GRAFT.

N

Patient 1

THE MICROSCOPE WAS INSPECTED BY A SERVICE TECH ON APRIL 29, 2008 AND FOUND TO BE IN PROPER WORKING ORDER. UPON MEETING THE MEMBERS OF THE HOSPITAL, IT WAS DETERMINED THAT THE USER OF THE MICROSCOPE HAD NOT ATTENDED THE IN-SERVICE WHEN THE MICROSCOPE HAD BEEN INSTALLED AND THAT THE USER WAS NOT FAMILIAR WITH THE XENON ILLUMINATION. AN ADD'L IN-SERVICE WAS OFFERED TO THE CUSTOMER.