COMFORT COOL THUMB CMC ABDUCTION SPLINT NC79498

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-05-30 for COMFORT COOL THUMB CMC ABDUCTION SPLINT NC79498 manufactured by North Coast Medical, Inc..

Event Text Entries

[873447] An occupational therapist at a hospital received a thumb cmc abduction splint as a sample from north coast medical, inc to try for use approval. She contacted north coast medical, inc in 2008 to report an incident with the splint. She read the instructions for heating the splint to soften the thermoplastic insert. The instructions say to heat the splint in 20 seconds intervals, not to exceed two minutes total heating time. She said she had scanned the instructions, and the 20 second intervals did not stand out. After the splint had been heating for a continuous one minute and 30 seconds, she smelled something burning and the splint was on fire. On removing the splint from the microwave she received a burn on her finger. She reported to the emergency room at her place of employment for treatment for the burn which was reported on the emergency room report as a first degree burn. She was immediately released from the emergency room after treatment, and is not required for follow up treatments. When the therapist contacted north coast medical she told her contact she received third degree burns. This contradicts the emergency room report. Therapist acknowledged the instructions were not followed for preparing the splint.
Patient Sequence No: 1, Text Type: D, B5


[8006160] This event occurred during the preparation of the splint prior to application. When the therapist noticed the splint was burning in the microwave, she reached in with her hand to remove it while it was still hot, which caused the first degree burn she suffered.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2939821-2008-00001
MDR Report Key1054634
Report Source05
Date Received2008-05-30
Date of Report2008-05-28
Date of Event2008-05-02
Date Mfgr Received2008-05-07
Date Added to Maude2009-11-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOCCUPATIONAL THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street18305 SUTTER BOULEVARD
Manufacturer CityMORGAN HILL CA 95037
Manufacturer CountryUS
Manufacturer Postal95037
Manufacturer Phone4087765000
Manufacturer G1SCOTT SPECIALTIES, INC.
Manufacturer Street512 M STREET
Manufacturer CityBELLEVILLE KS 69935
Manufacturer CountryUS
Manufacturer Postal Code69935
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOMFORT COOL THUMB CMC ABDUCTION SPLINT
Generic NameABDUCTION SPLINT
Product CodeFYH
Date Received2008-05-30
Returned To Mfg2008-05-19
Model NumberNC79498
Catalog NumberNC79498
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerNORTH COAST MEDICAL, INC.
Manufacturer AddressMORGAN HILL CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-05-30

© 2026 FDA.report