THE SPANNER TEMPORARY PROSTATIC STENT SPNR SPNR-7C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2008-05-30 for THE SPANNER TEMPORARY PROSTATIC STENT SPNR SPNR-7C manufactured by Abbeymoor Medical Inc..

Event Text Entries

[863066] During abbeymoor medical review of a physician provided cystoscopy video, the device manufacturer observed a complete device migration. The device was initially positioned in 2008, and removed by cystoscopy on approximately 2008. The physician stated the removal was simple and the device came out well. Indication for use is post acute urinary retention.
Patient Sequence No: 1, Text Type: D, B5


[8020187] The device was not returned for evaluation. A review of device history records (dhr) for the relevant spanner lot revealed no indication that the device used did not meet specifications. Physician feedback indicates the patient was able to urinate at the time of device removal. When attempting device removal, the physician couldn't find the retrieval suture and discovered the device migrated into the bladder. A cystoscopy was performed. This form fda 3500a is being submitted after the required 30-day time frame in response to recent interactions between abbeymoor, the office, and the fda reporting systems monitoring branch in which migration events with cystoscopic removal were deemed mdr reportable. (abbeymoor initially determined that an mdr was not required for this event based on results of a clinician-based risk assessment).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005249627-2008-00008
MDR Report Key1056485
Report Source05,07
Date Received2008-05-30
Date of Report2008-05-29
Date of Event2008-03-13
Date Mfgr Received2008-03-24
Device Manufacturer Date2007-07-01
Date Added to Maude2008-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street501 E SOO STREET
Manufacturer CityPARKERS PRAIRIE MN 56361
Manufacturer CountryUS
Manufacturer Postal56361
Manufacturer Phone2183386700
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHE SPANNER TEMPORARY PROSTATIC STENT
Generic NameTEMPORARY PROSTATIC STENT
Product CodeNZC
Date Received2008-05-30
Model NumberSPNR
Catalog NumberSPNR-7C
Lot Number159
ID Number2009004-7C
Device Expiration Date2008-12-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1267574
ManufacturerABBEYMOOR MEDICAL INC.
Manufacturer Address501 E SOO STREET PARKERS PRAIRIE MN 56361 US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2008-05-30

© 2026 FDA.report