BACT/ALERT MP CULTURE BOTTLES 259797

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2008-06-25 for BACT/ALERT MP CULTURE BOTTLES 259797 manufactured by Biomerieux, Inc..

Event Text Entries

[882585] A plastic bact/alert mp culture bottle was found broken in the bact/alert incubator. Afb stain indicated that its contents were negative. There were no adverse events noted as a result of this incident. The lab was closed and a formaldehyde gas sterilization was performed.
Patient Sequence No: 1, Text Type: D, B5


[8020326] The user had disinfected the plastic bact/alert mp culture bottle with biocleanse (teknom), a disinfectant, which according to teknon's application chart, is not recommended for the use on polycarbonate bottles. Biomerieux, inc. Has investigated the interaction of some disinfectants on its polycarbonate bottles and has issued a notice alerting users to read the disinfectant's label to determine compatibility with polycarbonate bottles. Biomerieux has also modified its package insert with a caution about polycarbonates and choice of disinfectants by the user.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002769706-2008-00002
MDR Report Key1065518
Report Source06
Date Received2008-06-25
Date of Report2008-06-20
Date of Event2008-05-27
Date Mfgr Received2008-05-28
Device Manufacturer Date2007-11-01
Date Added to Maude2008-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactELISABETH FRITZ
Manufacturer Street100 RODOLPHE ST
Manufacturer CityDURHAM NC 27712
Manufacturer CountryUS
Manufacturer Postal27712
Manufacturer Phone9196202682
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBACT/ALERT MP CULTURE BOTTLES
Generic NameMICROBIAL GROWTH MONITOR
Product CodeJTA
Date Received2008-06-25
Catalog Number259797
Lot Number1018685
Device Expiration Date2008-11-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1150017
ManufacturerBIOMERIEUX, INC.
Manufacturer Address100 RODOLPHE ST DURHAM NC US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-06-25

© 2024 FDA.report