[889899] It was reported that during surgery while using the illumination system, the cable melted, creating smoke/heat and would not function properly. No patient complications were reported.
Patient Sequence No: 1, Text Type: D, B5


[7891053] The light cable was returned to medtronic for evaluation. A visual examination of the cable confirms high temps present at the light source attachment end. This suggests the use of a machine with greater than the 100w instructed on the packaging. Package labeling indicates that the recommended light source utilizes 100w light sources and 5mm optic cables. Use of larger cables and/or higher wattage light sources may result in high temperatures on the metal connection to the light cable, which may result in injury to patient or staff and damage to product.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1030489-2008-00308
MDR Report Key1070061
Report Source05
Date Received2008-07-03
Date of Report2008-06-06
Date of Event2008-05-23
Date Mfgr Received2008-06-06
Device Manufacturer Date2008-03-20
Date Added to Maude2008-10-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHAD ASHTON
Manufacturer Street1800 PYRAMID PLACE
Manufacturer CityMEMPHIS TN 38132
Manufacturer CountryUS
Manufacturer Postal38132
Manufacturer Phone9013963133
Manufacturer G1WARSAW ORTHOPEDIC INC
Manufacturer Street2500 SILVEUS CROSSING
Manufacturer CityWARSAW IN 46852
Manufacturer CountryUS
Manufacturer Postal Code46852
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMAST QUADRANT RETRACTOR SYSTEM
Generic NameILLUMINATION SYSTEM
Product CodeFSZ
Date Received2008-07-03
Returned To Mfg2008-06-09
Model NumberNA
Catalog Number9560658
Lot NumberW08B1264
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1208611
ManufacturerWARSAW ORTHOPEDIC INC.
Manufacturer Address2500 SILVEUS CROSSING WARSAW IN 46852 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-07-03