TYDINGS TONSIL SNARE, 7 3/8", 12 SNARE WIRES, 25 G 450-510

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2008-07-07 for TYDINGS TONSIL SNARE, 7 3/8", 12 SNARE WIRES, 25 G 450-510 manufactured by .

Event Text Entries

[21254977] The user facility reports that the tonsil snare failed to cut through tissue and release, necessitating the surgeon to remove additional tissue to free the snare and tonsil tissue. The snare push rod that is attached to the movable handle, jumped out of the channel and jammed, preventing the snare contraction around tonsil. The surgery technician had to remove the thumbscrew to reset the snare. It appears that the distal channel is bent. The physician was concerned about where the snare was positioned, too close to carotid artery.
Patient Sequence No: 1, Text Type: D, B5


[21433723] The device involved in the reported incident has been received for evaluation. An investigation has been initiated based on the reported information. Integra lifesciences corporation is filing on behalf of the initial distributor j. Jammer surgical instruments.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2430952-2008-00023
MDR Report Key1073942
Report Source05,06,07
Date Received2008-07-07
Date of Report2008-07-07
Date of Event2008-06-17
Date Facility Aware2008-06-17
Report Date2008-07-07
Date Mfgr Received2008-06-17
Date Added to Maude2010-04-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSANDRA LEE, RN
Manufacturer Street311C ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099366828
Manufacturer G1IIG INSTRUMENTE GMBH
Manufacturer StreetDORFSTRASSE 2
Manufacturer CityDURCHHAUSEN 78591
Manufacturer CountryGM
Manufacturer Postal Code78591
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTYDINGS TONSIL SNARE, 7 3/8", 12 SNARE WIRES, 25 G
Generic NameJARIT THROAT/LARYNGEAL
Product CodeKBZ
Date Received2008-07-07
Returned To Mfg2008-06-20
Catalog Number450-510
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2008-07-07