GX770 INTRAORAL X-RAY 46-404600G6 A0770JT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2008-07-10 for GX770 INTRAORAL X-RAY 46-404600G6 A0770JT manufactured by Gendex Dental Systems.

Event Text Entries

[881035] The tubehead of the gx770 i/o unit fell off of the yoke and landed on the pt's chest.
Patient Sequence No: 1, Text Type: D, B5


[8067876] The pt was unharmed from the incident. The tubehead came free from the yoke. A set screw was missing. Conclusions - when a replacement screw is sent and received, the unit will be repaired by the service technician.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004115000-2008-00013
MDR Report Key1075015
Report Source05
Date Received2008-07-10
Date of Report2008-06-13
Date of Event2008-06-04
Date Mfgr Received2008-06-13
Device Manufacturer Date2006-12-05
Date Added to Maude2009-11-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARI LAMBERT, MGR
Manufacturer Street901 WEST OAKTON
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone8473643958
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGX770 INTRAORAL X-RAY
Generic NameDENTAL X-RAY EQUIPMENT
Product CodeEAP
Date Received2008-07-10
Model Number46-404600G6
Catalog NumberA0770JT
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGENDEX DENTAL SYSTEMS
Manufacturer Address901 WEST OAKTON DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2008-07-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.