REDY 2000 N/A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1992-07-04 for REDY 2000 N/A manufactured by Organon Teknika.

Event Text Entries

[20437487] 4/28/92 floor nurse called maintenance-reported the redy 2000 had a malfunction. Bio-med technician responded and reported lowflow alarms during priming mode. Called manufacturer and reported malfunction. 4/29/92 organon-teknika field engineer came and replaced flow transducer and checked out system. Patient was an 85 year old white female who was hospitalized with chf and renal failure. Dialysis machine would not function out of priming mode. Patient ultimately transferred to another facility for dialysisdevice not labeled for single use. Patient medical status prior to event: critical condition. There was not multiple patient involvement. Device serviced in accordance with service schedule. Date last serviced: 01-apr-92. Service provided by: manufacturer. Service records available. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, mechanical tests performed, performance tests performed. Results of evaluation: component failure, transducer. Conclusion: invalid data. Certainty of device as cause of or contributor to event: invalid data. Corrective actions: device repaired and put back in service. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1077
MDR Report Key1077
Date Received1992-07-04
Date of Report1992-04-30
Date of Event1992-04-28
Date Facility Aware1992-04-28
Report Date1992-04-30
Date Reported to FDA1992-04-30
Date Reported to Mfgr1992-04-28
Date Added to Maude1992-08-18
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameREDY 2000
Generic NameDIALYSIS SYSTEM
Product CodeFLC
Date Received1992-07-04
Model Number2000
Catalog NumberN/A
Lot NumberN/A
ID NumberCRH #3518
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Age01-MAR-92
Implant FlagN
Device Sequence No1
Device Event Key1033
ManufacturerORGANON TEKNIKA


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1992-07-04