OUTPATIENT II OPII NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2008-05-20 for OUTPATIENT II OPII NA manufactured by Burton Medical Products.

Event Text Entries

[881607] Light fell at extension arm junction during procedure. No injuries occurred.
Patient Sequence No: 1, Text Type: D, B5


[7885187] The light was in use when it separated from the down tube at the extension arm junction. No injuries occurred and the light was replaced. The unit shows signs of excessive wear and tear. The arm was continually fatigued by the end-user pushing the extension arm up into the pivot support. This caused the extension arm to separate and drop. The end-suer should have noticed the damage occurring at the junction and had the light repaired prior to the incident.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2018492-2007-00026
MDR Report Key1077323
Report Source06
Date Received2008-05-20
Date of Report2007-11-19
Date of Event2007-10-25
Date Mfgr Received2007-10-25
Device Manufacturer Date2005-02-01
Date Added to Maude2010-04-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHELSEA MITCHELL
Manufacturer Street21100 LASSEN ST.
Manufacturer CityCHATSWORTH CA 91311
Manufacturer CountryUS
Manufacturer Postal91311
Manufacturer Phone8187018700
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOUTPATIENT II
Generic NameDIAGNOSTIC LIGHT
Product CodeFCQ
Date Received2008-05-20
Model NumberOPII
Catalog NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBURTON MEDICAL PRODUCTS
Manufacturer AddressCHATSWORTH CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-05-20

© 2026 FDA.report