MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1994-01-05 for AIR-VAC TRANSPORT INCUBATOR manufactured by Ohmeda.
[17442064]
Inside the incubator's plexiglass hood, there is a thermometer in a thermometer holder which is glued to the hood. The holder becomes ungluded and falls into the incubator. Recently, there were 2 incidents where the thermometer barely missed infants. The reporter notified the mfr of the probelm as far back as 3 yrs ago. Since 1987, there have been 21 incidents; 4 reports were made to the mfr. This info was submitted as f/u 3/10/94: another incident occured on 3/4/94.
Patient Sequence No: 1, Text Type: D, B5
Report Number | MW1000299 |
MDR Report Key | 10790 |
Date Received | 1994-01-05 |
Date of Report | 1994-01-04 |
Date Added to Maude | 1994-01-06 |
Event Key | 0 |
Report Source Code | Voluntary report |
Manufacturer Link | N |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Reporter Occupation | BIOMEDICAL ENGINEER |
Health Professional | 3 |
Initial Report to FDA | 0 |
Report to FDA | 0 |
Event Location | 3 |
Single Use | 0 |
Previous Use Code | 0 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | AIR-VAC TRANSPORT INCUBATOR |
Product Code | FPL |
Date Received | 1994-01-05 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Implant Flag | N |
Date Removed | B |
Device Sequence No | 1 |
Device Event Key | 10790 |
Manufacturer | OHMEDA |
Manufacturer Address | MADISON WI 537077550 US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1994-01-05 |