[903562] Brand new unit was undergoing initial acceptance testing in the biomedical department. The unit was visually inspected and showed no signs of damage and was in new condition. The unit was filled with water and connected to an electrical safety analyzer for the initial electrical safety test. Upon powering up the unit, the electrical safety analyzer began smoking and the 20 amp electrical breaker tripped in the electrical distribution panel. The manufacturer was notified and a replacement hydrocollator was sent. It is obvious that the unit is miswired and was a serious safety hazard. The electrical safety analyzer was destroyed. ====================== manufacturer response for hot pack machine, m4 hydrocollator======================tech support notified and a replacement unit was sent. No response from regulatory affairs on whether or not our safety analyzer will be repaired/replaced.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1082891
MDR Report Key1082891
Date Received2008-07-18
Date of Report2008-07-18
Date of Event2008-07-10
Report Date2008-07-18
Date Reported to FDA2008-07-18
Date Added to Maude2008-07-30
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameHYDROCOLLATOR
Product CodeIRQ
Date Received2008-07-18
Returned To Mfg2008-07-18
Model NumberM-4
Catalog Number*
Lot Number*
ID Number*
OperatorOTHER
Device AvailabilityR
Device Age1 DY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key1048610
ManufacturerCHATTANOOGA GROUP
Manufacturer Address4717 ADAMS STREET HIXSON TN 37343 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-07-18