MEDICHOICE UMBILICAL CORD CLAMP 1773088101 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2008-08-14 for MEDICHOICE UMBILICAL CORD CLAMP 1773088101 NA manufactured by Changzhou Concorde Plastic Products Co., Ltd.

Event Text Entries

[895078] A problem was brought to our attention with the item 1773088101 umbilical cord clamp. The clamp was used on an infant in conjunction with another device called the "my child infant protection system". This device is a transponder that is attached to the hinge of the umbilical cord and acts as a security device in the event of child abduction from a hospital setting. The baby was nursing on the mother when a 20 minute check was performed. Once the baby was removed from the mother, the nurse noticed that the blanket was soaked with approximately 50cc of blood. The amount of blood was determined by weighing the wet blanket against a dry blanket. The nurse also noticed that the cord clamp was not closed and locked properly, because the umbilical cord was large with the artery on the side. Both the mother and child are diabetic. The baby was moved to the neonatal icu to monitor the blood sugar levels after the incident. The baby was subsequently discharged from the hospital. As a result of this incident, the hospital has instructed its nurses to use two clamps with hinges opposed to large cords.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1451040-2008-00001
MDR Report Key1114423
Report Source99
Date Received2008-08-14
Date of Report2008-08-14
Date of Event2008-07-16
Date Facility Aware2008-07-21
Report Date2008-08-14
Date Reported to FDA2008-08-14
Date Reported to Mfgr2008-08-14
Date Added to Maude2009-11-18
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMEDICHOICE UMBILICAL CORD CLAMP
Generic NameUMBILICAL CORD CLAMP
Product CodeHFW
Date Received2008-08-14
Model Number1773088101
Catalog NumberNA
Lot NumberSG08-1
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age11 MO
Device Sequence No1
Device Event Key0
ManufacturerCHANGZHOU CONCORDE PLASTIC PRODUCTS CO., LTD
Manufacturer AddressHUADU INDUSTRY AREA QUINHUANG TOWN WUJIN, CHANGZHOU, JINGSU 213173 CH 213173


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-08-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.