* LP-800B

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1997-08-14 for * LP-800B manufactured by Aadco Advanced Products.

Event Text Entries

[71072] This letter is to inform dist that, while surveying three 0. 5 mm pb thyroid shields facility has measured radiation exposure readings which exceed normal radiation exposure readings. The following table summarizes the results of radiation surveys: geiger-muller survey instrument: ludlum 14 c, background exposure reading=0. 03 mr/hr. Thyroid shield (a), lp-800b, mfr data 5/97 yielded 0. 12 mr/hr. Thyroid shield (b), lp-8015, mfr date 5/97, yielded 0/13 mr/hr, and thyroid shield (c), lp-800b, mfr date 5/97, yielded 0/14 mr/hr.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1011883
MDR Report Key113177
Date Received1997-08-14
Date of Report1997-08-04
Date Added to Maude1997-08-18
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationPHYSICIST
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand Name*
Generic NameTHYROID SHIELD A, & C
Product CodeKPY
Date Received1997-08-14
Model Number*
Catalog NumberLP-800B
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key111106
ManufacturerAADCO ADVANCED PRODUCTS
Manufacturer AddressRT 66, P.O. BOX 410 CATAMOUNT PARK RANDOLPH VT 05060 US

Device Sequence Number: 2

Brand Name*
Generic NameTHYROID SHIELD B
Product CodeKPY
Date Received1997-08-14
Model Number*
Catalog NumberLP-8015
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Age*
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key111109
ManufacturerAADCO ADVANCED PRODUCTS
Manufacturer AddressRT 66, P.O. BOX 410 CATAMOUNT PARK RANDOLPH VT 05060 US