ARTHRO-KNIFE, SERRATED STRAIGHT 2-SIDED ROUND 70-0078-103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06 report with the FDA on 2008-08-19 for ARTHRO-KNIFE, SERRATED STRAIGHT 2-SIDED ROUND 70-0078-103 manufactured by Conmed Linvatec.

Event Text Entries

[940437] The customer reported that during pre-operative testing of this device for a pylormyotomy procedure, the blade did not fully retract into the protective sheath. This event occurred with pre-testing of two knives. A third arthro-knife was obtained and functioned during testing. The third knife was used to complete the procedure. There was no injury or significant surgical delay reported with this event.
Patient Sequence No: 1, Text Type: D, B5


[8192676] Investigation results: conmed linvatec received this artho-knife for evaluation and the reported event was confirmed. The evaluation found that the end of the knife does not fully retract into the sheath. A corrective action is in process to address this failure mode. Conmed linvatec will continue to monitor this device for failures. Associated mdr: 1017294-2008-00292.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017294-2008-00293
MDR Report Key1135817
Report Source00,06
Date Received2008-08-19
Date of Report2008-07-30
Date of Event2008-07-30
Date Mfgr Received2008-07-30
Device Manufacturer Date2008-02-01
Date Added to Maude2010-04-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactBEVERLY SCHANER
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal33773
Manufacturer Phone7273995169
Manufacturer G1CONMED LINVATEC
Manufacturer Street11311 CONCEPT BLVD.
Manufacturer CityLARGO FL 33773
Manufacturer CountryUS
Manufacturer Postal Code33773
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARTHRO-KNIFE, SERRATED STRAIGHT 2-SIDED ROUND
Generic NameARTHRO-CUT KNIFE
Product CodeGAC
Date Received2008-08-19
Returned To Mfg2008-08-06
Model NumberNA
Catalog Number70-0078-103
Lot NumberBBD59412
ID NumberNA
Device Expiration Date2013-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCONMED LINVATEC
Manufacturer AddressLARGO FL US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-08-19

© 2026 FDA.report