CRYOCYTE FREEZING CONTAINER 500 ML W/LABEL POCKET R4R9955

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2008-08-22 for CRYOCYTE FREEZING CONTAINER 500 ML W/LABEL POCKET R4R9955 manufactured by Baxter Healthcare Corporation - Mountain Home.

Event Text Entries

[21992543] (b) (4).
Patient Sequence No: 1, Text Type: N, H10


[22126989] This is a report received on 05-august-2008 by baxter (b) (4) via the distributor, (b) (4) from a healthcare professional. A leakage (crack at the upper corner) on a cryocyte bag was detected after thawing. The cells were transfused from the over-pouch. The product has been used before in the same setting frequently. Additional info received indicates that the total number of target stem cells transplanted was 2. 8x10e6/kg. The intended patient experienced delayed transfusion, prophylactic antibiotics, and a sterility risk. Per the customer's cryocyte processing protocol: dmso was used as a cryopreservant. Any visible air was removed from the bag and an over-wrap was used as well as a controlled rate freezer. The bag is then stored in vapor phase nitrogen for 30 days. The bag was not removed from the final storage temperature at any time. For transport, the bag was placed in a dry shipper filled with liquid nitrogen. The customer was not aware of any damage or deviation from standard procedure that may have occurred to the bag during filling, freezing, storage, transport, or thawing. In addition, there have been no recent changes in protocol, critical equipments/supplies or personnel. No additional information is available.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423500-2008-00763
MDR Report Key1137533
Report Source01,05
Date Received2008-08-22
Date of Report2008-08-05
Date of Event2008-07-28
Date Mfgr Received2008-08-05
Date Added to Maude2009-12-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKAREN KIRBY, MANAGER
Manufacturer Street25212 W. ILLINOIS ROUTE 120
Manufacturer CityROUND LAKE IL 60073
Manufacturer CountryUS
Manufacturer Postal60073
Manufacturer Phone8472704541
Manufacturer G1BAXTER HEALTHCARE CORPORATION
Manufacturer Street1900 HIGHWAY 201 NORTH
Manufacturer CityMOUNTAIN HOME AR 72653249
Manufacturer CountryUS
Manufacturer Postal Code72653 2497
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCRYOCYTE FREEZING CONTAINER 500 ML W/LABEL POCKET
Generic Name81KSE
Product CodeKSE
Date Received2008-08-22
Catalog NumberR4R9955
Lot NumberH06B01040
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE CORPORATION - MOUNTAIN HOME
Manufacturer AddressMOUNTAIN HOME AR US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-08-22

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.