RPM RESPIRATORY GATING H50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2008-08-28 for RPM RESPIRATORY GATING H50 manufactured by Varian Medical Systems.

Event Text Entries

[947004] The customer reported, "a plan is created on the ct dataset of the pt in a certain respiratory phase (deep inspiration breath hold in our case). The toggle "use gated" in the general tab of the plan properties (in eclipse or even in rt-chart) is on. The plan is loaded in the 4d itc. If the key of the rpm system (the box enabling the usage of the gating system) is in the position gating disabled, the plan is treatable! In other words, there is no interlock preventing the irradiation of the plan supposed with gating (as per the toggle "use gated"). A pt was mis-administered. "
Patient Sequence No: 1, Text Type: D, B5


[8193202] Other: based on the initial report, varian's medical advisor was not able to render a medical evaluation on whether this event may have caused or contributed to a serious injury, or would be likely to cause or contribute to a death or serious injury if the malfunction were to recur. Though still under investigation, varian has determined that a mdr is appropriate. Additional follow-up to this mdr is expected upon completion of the investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916710-2008-00039
MDR Report Key1139550
Report Source05,06
Date Received2008-08-28
Date of Report2008-07-21
Date of Event2008-07-10
Date Mfgr Received2008-07-21
Device Manufacturer Date2006-07-01
Date Added to Maude2008-09-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVE HALL
Manufacturer Street911 HANSEN WAY
Manufacturer CityPALO ALTO CA 94304
Manufacturer CountryUS
Manufacturer Postal94304
Manufacturer Phone6504833153
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRPM RESPIRATORY GATING
Generic NameECG/RESPIRATOR
Product CodeIXO
Date Received2008-08-28
Model NumberH50
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrN
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1156238
ManufacturerVARIAN MEDICAL SYSTEMS
Manufacturer AddressPALO ALTO CA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-08-28

© 2024 FDA.report