PANBIO WEST NILE VIRUS IGM CAPTURE ELISA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-09-09 for PANBIO WEST NILE VIRUS IGM CAPTURE ELISA manufactured by Inverness Medical Innovations Australia Pty Ltd 532.

Event Text Entries

[920597] Statement of problem: about 80% of the pt specimens collected in the month of 2008 that were reported by viromed as positive for west nile igm antibody using the panbio west nile virus igm capture elisa, failed to confirm by a center panel of serologic test results including igm capture elisa, microsphere immunoassay to west nile envelope and ns5, and immunofluorescence assay to encephalitis virus. The probable source of error is the mfr's selection of assay cutoff calibrator. Viromed has confirmed independently an assay false positive rate of about 80%. Viromed is a national resource to physicians for west nile virus serology. Pt care providers, and public health agencies should consider as suspect positive reports for west nile virus igm antibody where the panbio west nile virus igm capture elisa was used for specimen analysis.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5008254
MDR Report Key1156004
Date Received2008-09-09
Date of Report2008-09-09
Date Added to Maude2008-09-18
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NamePANBIO WEST NILE VIRUS IGM CAPTURE ELISA
Generic NamePANBIO WEST NILE VIRUS IGM CAPTURE ELISA
Product CodeLRM
Date Received2008-09-09
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1183119
ManufacturerINVERNESS MEDICAL INNOVATIONS AUSTRALIA PTY LTD 532
Manufacturer AddressSEVENTEEN MILE ROCKS RD. SINNAMON PARK QUEENSLAND AS 4073


Patients

Patient NumberTreatmentOutcomeDate
10 2008-09-09