* TS-2 R - WITH RING ATTACHMENT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1997-08-21 for * TS-2 R - WITH RING ATTACHMENT manufactured by Burlington Medical Supplies.

Event Text Entries

[71863] In response to the advisory, all leaded radiation protection devices were evaluated for radioactive contamination using a ludlum 14 c survey meter with a ludlum model 44-9 pancake probe. One contaminated device was found. The survey results are. 5 mr/hr at the surface. The supplier will be contacted by phone to discuss a replacement device and return shipping arrangements.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1011961
MDR Report Key115767
Date Received1997-08-21
Date Added to Maude1997-08-28
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand Name*
Generic NameTHYROID PROTECTIVE COLLAR
Product CodeKPY
Date Received1997-08-21
Model NumberTS-2 R - WITH RING ATTACHMENT
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device Availability*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key113597
ManufacturerBURLINGTON MEDICAL SUPPLIES
Manufacturer Address3 ELMHURST ST. NEWPORT VA 23603 US


Patients

Patient NumberTreatmentOutcomeDate
10 1997-08-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.