M2000SP 9K014-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-09-15 for M2000SP 9K014-01 manufactured by Abbott Molecular, Inc..

Event Text Entries

[17397120] The front hood panel of the m2000sp instrument has adhesive on the double-coated tape that was not strong enough to hold the window in place. Five internal instances have been documented. Additionally, five customer observations of the hood front window falling off have been documented. None of these individual instances have resulted in a medical device report (form 3500a). However, abbott molecular (am) took a field action in 2008. Thus abbott molecular is filing this mdr due to this situation's malfunction and abbott molecular's field correction.
Patient Sequence No: 1, Text Type: D, B5


[17638453] Abbott molecular has identified a replacement tape to secure the front hood panel. A customer letter has been issued informing the customers that have a hood on their m2000sp system that a service call will be scheduled. If the customer suspects that the front hood panel is loose, then the m2000sp instrument is not to be used until the service visit is complete. Additional model #: 9k14-02.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005248192-2008-00002
MDR Report Key1161496
Report Source07
Date Received2008-09-15
Date of Report2008-09-12
Date of Event2008-08-25
Date Mfgr Received2008-08-25
Date Added to Maude2009-08-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPAULA MARTIN, SENIOR MGR
Manufacturer Street1300 EAST TOUHY AVENUE
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2243617333
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameM2000SP
Generic NameAUTOMATED SAMPLE PREPARATION SYSTEM
Product CodeKEY
Date Received2008-09-15
Model Number9K014-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer AddressDES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-09-15