ICON DS STREP A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-09-16 for ICON DS STREP A manufactured by Beckman Coulter.

Event Text Entries

[19177340] Beckman coulter icon ds strep a. This is a point-of-care rapid test to diagnose strep throat. The printing on the test strips is upside-down as compared to the procedure card -in the box- examples. If someone were to hold the strip in a way that allowed them to read the words "control" and "test" while trying to follow the card instructions, they could read a negative test as positive, and vice versa. Diagnosis or reason for use: rapid strep test for diagnosis of strep throat.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5008325
MDR Report Key1164224
Date Received2008-09-16
Date of Report2008-09-16
Date Added to Maude2008-09-24
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameICON DS STREP A
Generic NameNONE
Product CodeGTY
Date Received2008-09-16
Lot Number426H11
Device Expiration Date2008-02-29
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1193358
ManufacturerBECKMAN COULTER


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-09-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.