CAPTURE-R SELECT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2008-09-24 for CAPTURE-R SELECT manufactured by Immucor, Inc..

Event Text Entries

[963964] During galileo validation, unexpected negative reactions were observed using capture-r select. Eighty donors were tested and half of them demonstrated positive reactivity when tested with a serum know to contain anti-vel.
Patient Sequence No: 1, Text Type: D, B5


[8193998] The customer returned a sample (reported to contain anti-vel), it was tested with panoscreen, lot 28317, using immuadd as the potentiator. Incubations at is and rt were included. No reactivity was observed at is or rt. Variable reactivity was observed at iat. The customer's sample was also tested with capture-r ready screen (crrs), lot g140, and capture-r select, lot sc050. The sample was reactive with cell iv on crrs. The sample was reactive with all cells on select, lot sc050. No vel-negative donor cells were available for testing. The customer did not return product for investigation, the results appear to be due to the nature of the sample.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1034569-2008-00472
MDR Report Key1171475
Report Source01,05,06
Date Received2008-09-24
Date of Report2008-09-24
Date of Event2008-09-01
Date Mfgr Received2008-09-01
Date Added to Maude2008-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. J. SCOTT WEBBER
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 30071
Manufacturer CountryUS
Manufacturer Postal30071
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 30071
Manufacturer CountryUS
Manufacturer Postal Code30071
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R SELECT
Generic NameSOLID PHASE RED CELL ADHERENCE ASSAY
Product CodeKSS
Date Received2008-09-24
Lot NumberSC050
Device Expiration Date2006-11-30
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrR
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1196768
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 30071 US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-09-24

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.