The customer stated that a large bias (22%) of bilirubin results was observed for pt results generated on the architect c8000 analyzer. The customer pulled 442 samples to rerun the tests, 80 pt results were changed after they were originally reported out as abnormal. Impact to pt management is unk.
Patient Sequence No: 1, Text Type: D, B5
[8217505]
Abbott customers reported the following issues for clinical chemistry bilirubin calibrator: low bias for total bilirubin quality control (qc) and pt results with bilirubin calibrator lot number change and results trending low with proficiency survey samples. Investigation of these issues determined that the calibrator values assigned to the total bilirubin reagent (list 6l45 and 8g62) caused results to trend downward over the past several calibrator lot numbers. On 08/10/2007 a product correction letter (fa10aug2007) was issued to address bilirubin low recovery complaints and provided customers with new bilirubin values for all on-market calibrator lots. The new bilirubin values were assigned using the secondary standard (b)(6). (b)(6) standard preparation was identified as a source of variability. The (b)(6) standard minimized lot-to-lot variability in the value assignment process. Following the use of (b)(6) as a standard for the assignment for the bilirubin calibrator, an increase in customer complaints for high results for total bilirubin reagents (06l45 and 08g62) was observed: higher than expected results on proficiency survey samples, higher than expected quality control (qc) recovery, and higher than expected pt results. An investigation was initiated for the total bilirubin reagents. The root cause was determined to be matrix issues for total bilirubin reagent (diazo) related to the commutability of various samples ((b)(6) standard, (b)(6) standard, bilirubin calibrator, and pt samples) due to the production materials used, e. G. Human serum albumin (hsa) vs. Bovine serum albumin (bsa). This matrix effect was responsible for the over-recovery observed by customers during the conversion to the (b)(6) standard. The cause of this issue was confirmed through testing performed using an external reference jendrassic and groff method. As a corrective/preventive measure all on-market bilirubin calibrator lots were re-value assigned per the new value assignment procedure providing improved accuracy for total bilirubin reagents. These values were communicated in the customer letter that was issued on 09/17/2008 (fa12sep2008 revision 01). In addition, abbott (b)(4) has re-written and validated the procedure for bilirubin calibrator. Value assignment and value verification that include: clarified procedure for primary (b)(6) standard preparation and increasing replicates to improving accuracy; revalidated the procedure. Reassigned of (b)(6) standards using in-house prepared (b)(6) standard to allow for less variability from one assignment to another. Assigned the values for bilirubin calibrator using reassigned (b)(6) standards by in-house prepared (b)(6) standards per jendrassic and groff method. This is the final report.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2018433-2008-00194 |
MDR Report Key | 1203491 |
Report Source | 05 |
Date Received | 2008-10-17 |
Date of Report | 2008-09-19 |
Date of Event | 2008-09-19 |
Date Mfgr Received | 2008-09-19 |
Device Manufacturer Date | 2007-11-01 |
Date Added to Maude | 2011-08-01 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | KHAUDEJA BANO, MD |
Manufacturer Street | 100 ABBOTT PARK RD DEPT 09Y6, AP6C-2 |
Manufacturer City | ABBOTT PARK IL 600646112 |
Manufacturer Country | US |
Manufacturer Postal | 600646112 |
Manufacturer Phone | 8479386319 |
Manufacturer G1 | ABBOTT MFG INC |
Manufacturer Street | 820 MISSION ST. |
Manufacturer City | SOUTH PASADENA CA 91030 |
Manufacturer Country | US |
Manufacturer Postal Code | 91030 |
Single Use | 3 |
Remedial Action | NO |
Previous Use Code | 3 |
Removal Correction Number | 2018433-09/19/08-004-C |
Event Type | 3 |
Type of Report | 3 |
Brand Name | CLINICAL CHEMISTRY BILIRUBIN CALIBRATOR |
Generic Name | FOR THE CALIBRATION OF CC BILI |
Product Code | JFM |
Date Received | 2008-10-17 |
Catalog Number | 1E66-04 |
Lot Number | 57919M100 |
Device Expiration Date | 2009-05-31 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MANUFACTURING INC |
Manufacturer Address | 820 MISSION ST. SOUTH PASADENA CA 91030 US 91030 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2008-10-17 |