Event Text Entries

[21630473] Oxygen flow meter was found to be "plugged in" a nitrous oxide outlet. Flow was found to be available. There is no evidence to suggest patients have ever received nitrous oxide. There is no evidence to suggest that oxygen flow meter was ever used. Examination demonstrates connector has a break in metal and plastic component which allows connection to enter nitrous oxidedevice not labeled for single use. Patient medical status prior to event: invalid data. Invalid data - regarding multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. No imminent hazard to public health claimed. Device not used as labeled/indended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: design - human factors. Conclusion: none or unknown. Certainty of device as cause of or contributor to event: invalid data. Corrective actions: device permanently removed from service, other. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1205
MDR Report Key1205
Date Received1992-08-25
Date of Report1992-07-24
Date of Event1992-07-16
Date Facility Aware1992-07-16
Report Date1992-07-24
Date Reported to FDA1992-07-24
Date Reported to Mfgr1992-07-24
Date Added to Maude1992-09-02
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameUNKNOWN
Generic NameOXYGEN FLOW METER
Product CodeBXY
Date Received1992-08-25
Model NumberUNKNOWN
Catalog NumberUNKNOWN
Lot NumberUNKNOWN
ID Number2982135
OperatorOTHER
Device AvailabilityY
Implant FlagN
Device Sequence No1
Device Event Key1153
ManufacturerCHEMETRON MEDICAL PRODUCTS


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1992-08-25