PRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT G1463J

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2008-10-10 for PRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT G1463J manufactured by Smiths Medical Asd, Inc..

Event Text Entries

[978880] User reports blood leaked from the syringe tip and droplets entered the doctor's eye. It is unknown if the unit was examined before use. No treatment given that we have been notified of.
Patient Sequence No: 1, Text Type: D, B5


[8211566] Results evaluations: our investigation into this event is based on the digital photos supplied by smiths medical. The sample deformity appears to be consistent with mechanical damage caused by the syringe barrel coming into contact with another ridged object during assembly. Previously, there was a dryer misalignment that was identified and corrected prior to the receipt of this complaint. It is hypothesized this dryer misalignment would cause this type of defect. The mfg records, of the two possible lots, were reviewed, and found to have no notation regarding this issue. Manufacturing has been notified of the issue, and has a heightened awareness of this type of defect. A review of our complaints database reveals no other reports on the two possible device lot numbers given. This report has been logged for trending.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217052-2008-00099
MDR Report Key1205708
Report Source07
Date Received2008-10-10
Date of Report2008-09-18
Date of Event2008-09-17
Date Mfgr Received2008-09-18
Date Added to Maude2009-11-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTIMOTHY MATSON
Manufacturer Street10 BOWMAN DRIVE
Manufacturer CityKEENE NH 03431
Manufacturer CountryUS
Manufacturer Postal03431
Manufacturer Phone6033523812
Manufacturer G1SMITHS MEDICAL ASD, INC.
Manufacturer CityKEENE NH 03431
Manufacturer CountryUS
Manufacturer Postal Code03431
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRO-VENT PLUS ARTERIAL BLOOD SAMPLING KIT
Generic Name75 GJE - TRAY, BLOOD COLLECTION
Product CodeGJE
Date Received2008-10-10
Catalog NumberG1463J
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITHS MEDICAL ASD, INC.
Manufacturer Address10 BOWMAN DRIVE KEENE NH 03431 US 03431


Patients

Patient NumberTreatmentOutcomeDate
10 2008-10-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.