CAMPBELL VENTRICULAR NEEDLE, 4", 12G 45-1065

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 1997-09-19 for CAMPBELL VENTRICULAR NEEDLE, 4", 12G 45-1065 manufactured by Johnson & Johnson Professionals, Inc..

Event Text Entries

[7762144] The device was returned to jjpi, on 10/2/97 and forwarded to the new bedord mfg facility for analysis. The returned cannula was observed to be badly pitted around the nickel plated luer hub. Indicating the possiblity of exposure to harsh or corrosive chemicals. The area around the end of the tube where the tip was attached was examined throughly. A considerable amount of solder was present around the rim and on the inside diameter of the tube. It can not be determined whether or not there was enough solder material without the tip, and the tip was not returned. The product complaint file showed no other complaints reported on this product. Jjpi considers this file closed.
Patient Sequence No: 1, Text Type: N, H10


[18552957] The pt under went a ventriculostomy. The tip of the needle was discovered missing after the physician had made his pass into the pt's ventricle. It was not known if the tip was there before making the pass, but was assumed to have been intact. After the case was completed, the pt was taken for a ct scan. The ct scan was negative.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219655-1997-00145
MDR Report Key121531
Report Source05
Date Received1997-09-19
Date of Report1997-09-19
Date of Event1997-08-21
Date Mfgr Received1997-08-21
Date Added to Maude1997-09-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCAMPBELL VENTRICULAR NEEDLE, 4", 12G
Generic NameCANNULA, VENTRICLE
Product CodeHCD
Date Received1997-09-19
Returned To Mfg1997-10-02
Model NumberNA
Catalog Number45-1065
Lot NumberUNK
ID NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key119042
ManufacturerJOHNSON & JOHNSON PROFESSIONALS, INC.
Manufacturer AddressNEW BEDFORD INDUSTRIAL PARK NEW BEDFORD MA 02745 US
Baseline Brand NameCAMPBELL VENTRICULAR NEEDLE, 4", 12G
Baseline Generic NameCANNULA, VENTRICLE
Baseline Model NoNA
Baseline Catalog No45-1065
Baseline IDNI


Patients

Patient NumberTreatmentOutcomeDate
10 1997-09-19

© 2026 FDA.report