THERMO FISHER SCIENTIFIC BPL25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2008-10-30 for THERMO FISHER SCIENTIFIC BPL25 manufactured by Thermo Fisher Scientific, Headquarters.

Event Text Entries

[967119] Recurring problems with blood bank refrigerator door handles. Door handles (lock & strike assembly) are having to be replaced on a quarterly basis. The door handles appear to be of poor design and made with inferior materials that lead to their constant breakage. Prior to their complete failure, the handles also cause problems with the closure of the refrigerators' doors. When this occurs, the refrigerators exhibit increased storage temps and temp alarms related to the door not closing properly and sealing, and lead potentially to the loss of blood products by way of the increased (excessive) storage temps. We feel that if the door handle was of better design and made of better (e. G. , stronger, more durable) materials, this problems would not occur as often as it is now.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5008817
MDR Report Key1216534
Date Received2008-10-30
Date of Report2008-10-24
Date of Event2008-10-21
Date Added to Maude2008-11-06
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameTHERMO FISHER SCIENTIFIC
Generic NameBLOOD BANK REFRIGERATOR
Product CodeKSE
Date Received2008-10-30
Model NumberBPL25
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No1
Device Event Key1266818
ManufacturerTHERMO FISHER SCIENTIFIC, HEADQUARTERS
Manufacturer Address81 WYMAN ST WALTHAM MA 02454 US

Device Sequence Number: 2

Brand NameTHERMO FISHER SCIENTIFIC
Generic NameBLOOD BANK REFRIGERATOR
Product CodeKSE
Date Received2008-10-30
Model NumberBBR25
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No2
Device Event Key1266819
ManufacturerTHERMO FISHER SCIENTIFIC, HEADQUARTERS
Manufacturer Address81 WYMAN ST WALTHAM MA 02454 US

Device Sequence Number: 3

Brand NameTHERMO FISHER SCIENTIFIC
Generic NameBLOOD BANK REFRIGERATOR
Product CodeKSE
Date Received2008-10-30
Model NumberBBR55
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date RemovedB
Device Sequence No3
Device Event Key1266820
ManufacturerTHERMO FISHER SCIENTIFIC, HEADQUARTERS
Manufacturer Address81 WYMAN ST WALTHAM MA 02454 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2008-10-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.