CRYOCYTE FREEZING CONTAINER 750 ML W/LABEL POCKET R4R9957

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2008-10-31 for CRYOCYTE FREEZING CONTAINER 750 ML W/LABEL POCKET R4R9957 manufactured by Baxter Healthcare.

Event Text Entries

[16120176] Baxter medical director assessment: this is a cryocyte bag leakage that occurred during/after thawing. There is no information of any pt adverse outcomes at this time. The customer site has refused to provide any additional pt information. As we do not know if the pt was infused with product we are unable to assess the risk of a break in sterility and a resulting infection due to the product being exposed to the outside atmosphere. As in all cases of cryocyte bag breakages during/after thawing there is the potential of loss of engraftment or delay in engraftment with the loss of product. This puts the pt at greater risk, with a possibility of infection. As such, this breakage could potentially cause or contribute to an event. As the customer site has refused to provide any additional information there is no further investigation necessary. Cryocyte bag breakage is currently being addressed through capa.
Patient Sequence No: 1, Text Type: N, H10


[16170468] During the thawing of the bags at 37 degree celsius customer recognized that cryocyte bags were leaking. The bags split along the seam. Four bags were affected. The units were stored in liquid nitrogen for 15 days; 150ml of stem cells were being stored. No freezing press was used to remove air. The customer has refused to give any additional data surrounding the pt.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2954761-2008-00052
MDR Report Key1220094
Report Source01,06
Date Received2008-10-31
Date of Report2008-10-31
Date of Event2008-09-26
Date Mfgr Received2008-10-02
Date Added to Maude2009-07-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationUNKNOWN
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOSHUA LEACH, SENIOR SPECIALIST
Manufacturer StreetONE BAXTER WAY
Manufacturer CityWESTLAKE VILLAGE CA 91362
Manufacturer CountryUS
Manufacturer Postal91362
Manufacturer Phone8053723250
Manufacturer G1BAXTER HEALTHCARE
Manufacturer Street1900 HIGHWAY, 201 NORTH
Manufacturer CityMOUNTAIN HOME AR 72653
Manufacturer CountryUS
Manufacturer Postal Code72653
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCRYOCYTE FREEZING CONTAINER 750 ML W/LABEL POCKET
Generic Name81KSE
Product CodeKSE
Date Received2008-10-31
Catalog NumberR4R9957
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerBAXTER HEALTHCARE
Manufacturer AddressMOUNTAIN HOME AR US


Patients

Patient NumberTreatmentOutcomeDate
10 2008-10-31

© 2026 FDA.report